Gtr-V3 Gas Transmission Rate Tester Gas Permeability Tester

Product Details
Customization: Available
After-sales Service: on Line
Warranty: 1year
Gold Member Since 2024

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

Importers and Exporters
The supplier has import and export rights
High Repeat Buyers Choice
More than 50% of buyers repeatedly choose the supplier
In-stock Capacity
The supplier has In-stock capacity
Fast Delivery
The supplier can deliver the goods within 15 days
to see all verified strength labels (26)
  • Gtr-V3 Gas Transmission Rate Tester Gas Permeability Tester
  • Gtr-V3 Gas Transmission Rate Tester Gas Permeability Tester
  • Gtr-V3 Gas Transmission Rate Tester Gas Permeability Tester
  • Gtr-V3 Gas Transmission Rate Tester Gas Permeability Tester
  • Gtr-V3 Gas Transmission Rate Tester Gas Permeability Tester
  • Gtr-V3 Gas Transmission Rate Tester Gas Permeability Tester
Find Similar Products

Basic Info.

Model NO.
GTR-V3
Display
Digital
Control
Computer Control
Weight
0-100Kg
Power Source
AC220V
Model
Gtr-V3
Measurement Range(Film)
0.1~50.000cc/M2.24h.0.1MPa
Sample Quantity
3
Main Sensor Quantity
3
Test Mode
Single Chamber Independant
Vacuum Precision
0.01PA
Vacuum Degree
<10PA
Temperature Control Range
5°c~95°c
Humidity Control Range
0%Rh
Temperature Control Precision
±0.05ºC
Temperature Resolution
0.01ºC
Size
760 mm (L) *575 mm (W) *450 mm (H)
Sample Thickness
<3mm,=3mm
Vacuum Tube Diameter
1/4"
Interface of Carrier Gas
Ф6
Transport Package
Wooden Packaging
Specification
GTR-V3
Trademark
paratronix
Origin
Jinan Shandong, China
HS Code
9024800000
Production Capacity
10000tower/Year

Product Description

It is suitable for the detection of O2, CO2, N2 and air permeability of plastic film, composite film, high barrier material, sheet, metal foil, rubber and other materials.


Technical parameters
Item Parameter
Model GTR-V3
Measurement range(film) 0.1~50.000cc/m2·24hours·0.1MPa
Sample quantity 3
Main sensor quantity 3
Test mode single chamber independant
Vacuum precision 0.01Pa
Vacuum degree <10Pa
Temperature control range 5ºC~95ºC
Humidity control range 0%RH,2%RH~98.5%RH±3%RH(note1)
Temperature control precision ±0.05ºC
Temperature resolution 0.01ºC
Test pressure -0.1 MPa~+0.3MPa(regular)
Air source pressure 0.4 MPa~0.6 MPa
Test air O2, N2 and CO2 (Outside of Supply)
Interface of carrier gas Ф6
Vacuum tube diameter 1/4"
Calibration mode Calibration of standard film and external calibration of vacuum sensor
Sample thickness ≤3mm
Sample size Φ97mm
Permeability area 38.48 cm2
Size 760 mm (L) × 575 mm (W) × 450 mm (H)

Gtr-V3 Gas Transmission Rate Tester Gas Permeability Tester
Test principles and product description
GTR-V3 adopts the differential pressure method test principle; the pre-processed sample is placed between the upper and lower test chambers, and clamped.First, the low pressure chamber (lower chamber) is vacuumized, and then the whole system is vacuumized; when reaching the specified vacuum degree, close the test chamber, to the high pressure chamber (upper chamber) filled with a certain pressure of the test gas, and ensure the formation of a constant pressure difference on both sides of the sample (adjustable); in this way, under the action of differential pressure gradient, the gas will be penetrated by the high pressure side and low pressure side. Through the monitoring and processing of the pressure inside the low pressure side, the barrier parameters of the test samples can be obtained. It fully complies with GB1038, ASTM D1434 and other standards.

Reference standards
YBB 00082003, GB/T 1038, GB/T7755.1, ASTM D1434,ISO 2556,ISO 15105-1,JIS K7126-A
Gtr-V3 Gas Transmission Rate Tester Gas Permeability Tester
Configuration
Standard configuration: host, constant temperature controller, computer, professional software, special sampler, vacuum grease, fast quantitative filter paper, vacuum pump
Optional: sampling blade, vacuum grease, vacuum pump oil, fast quantitative filter paper, humidity generator
Note: The air source of the machine is ф 6mm polyurethane pipe; gas source and distilled water are provided by users
The device appearance diagram shows the standard device appearance, the appearance of customized devices may vary.


Gtr-V3 Gas Transmission Rate Tester Gas Permeability Tester
Gtr-V3 Gas Transmission Rate Tester Gas Permeability Tester
Gtr-V3 Gas Transmission Rate Tester Gas Permeability Tester
FAQ

Are you a manufacturer or trading company?
We are the manufacturer and trading combo, which specialize in material testing instruments such as plastic package, medical device, facial mask, protective clothing and pharmaceutical package, etc. We located in Jinan, Shandong, China.

Do you have any certificate?
We  have 22 utility model patents and 17 software copyright certificates.

How about the installation?
The user's manual specifies the installation requirements with details. We will advise the relative information in advance, such as power supply, space, required temperature, humidity, etc.

Do you have after-sales service?
We will carry out the equipment inspection before delivery. If the machine does not work properly, we can make a remote control service to help dealing with the failure. We promise to reply within 24 hours and provide solutions within 2 working days.

How about the warranty?
Our warranty is 12 month, lifelong free of charge! User just bear the materials cost and express charge out-of-warranty period.

How long does the delivery time?
For the regular machine, 3-5 working days. For the customized one, 30 days or so.


 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now
Contact Supplier