Customization: | Available |
---|---|
After-sales Service: | on Line |
Warranty: | 12 Months |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Equipment type |
PMT-100N |
Device name | Needle tube rigidity toughness tester |
Equipment power supply |
220V 50Hz |
Overall dimension |
420(L)×350(W)×620(H) |
Equipment weight | About 30kg |
Rigid parameter | |
Measuring range | 100N |
Resolution | 0.01N |
Measurement accuracy | 0.3%F.S. |
Effective stroke | 150mm |
Support span | 1~60mm |
Deflection accuracy | 0.01mm |
Test speed | 0.1~200mm/mininfinitely variable speeds |
Toughness parameter | |
Sample size | 0.2~3.4mm |
Test frequency | 0.1~1.0Hz |
Bending Angle | 1~30° |
Bending distance | 5~100mm |
Range accuracy | ±0.1mm |
Angular accuracy | ±0.1° |
Number of bending |
1~999 times |
Rigidity: A specified force is applied to the center of the specified span of the needle being supported at both ends, and the deflection value of the needle is measured.
Toughness: Fix one end of the needle, apply a force from the fixed point to the specified span of the needle, first bend a specified Angle in one direction, and then bend a specified Angle in the opposite direction, and so repeatedly bend the specified number of times.
Standards
GB 15811-2016 Disposable Sterile Injection Needle
"YBB00112004-2015 Pre-potting Syringe Assembly (with needle)"
"YBB000992004-2015 Stainless Steel injection needle for pre-potting syringes"
GB/T 18457-2015 Stainless Steel syringes for manufacturing medical Devices